# CBCT artefacts: streaks near metal and motion blur

> Why a CBCT scan can show streaks next to metal and blur from movement, how often scans are repeated for motion (2 to 5%), and what it means for measuring.

- URL: https://jawviewer.com/blog/cbct-artefacts-streaks-near-metal-motion
- Category: Understanding your scan
- Published: 2026-11-21
- Updated: 2026-11-21
- Author: Jaw Viewer
- Reading time: 4 min
- Tags: cbct, image-quality, measurements

## Key points

- Artefacts are common in CBCT. They arise because the maths used to rebuild the scan does not match the real physics of the X-ray beam and the patient.
- In a bench test, titanium transmitted about 200% of what zirconia did at 60 kVp and about 530% at 90 kVp, and titanium streaks fell with higher beam energy while zirconia streaks did not.
- Repeat scans because of patient movement were 5.4% of 750 patients in an Italian study and 1.96% of 6364 in a Turkish one, highest in older adults and in children under 10.

An artefact is something in the picture that is not in the person. On a CBCT scan the two you meet most are streaks beside metal and blur from movement, and both come from how the scan is made. A 2011 review states that artefacts are common in cone beam CT and that they arise from discrepancies between the mathematical model used to rebuild the images and the actual physical imaging process.[1]

Metal and movement affect the scan differently. For the wider limits of the image, see [what a CBCT scan shows](https://jawviewer.com/blog/what-a-cbct-scan-shows.md).

## Why do artefacts exist at all?

A CBCT scanner takes many flat X-ray pictures as it turns around the head, and software rebuilds slices from them. The rebuilding assumes a tidy beam and a head that stays still. The same review notes that all the artefacts it discusses are more or less directly related to that reconstruction step, and that they can interfere with the diagnostic reading of a scan.[1]

So an artefact has a cause in the physics of the scan. A streak that crosses from one crown to the next is a product of the rebuild and has no tooth or bone behind it.

## What does different metal do?

Beam hardening from highly dense material such as metal is described as a common quality issue in jaw CBCT.[2] Not all metal behaves alike. A bench study put one zirconia implant (4.1 mm across) and one pure titanium rod (4.0 mm) in a CBCT unit and measured how much X-ray got through at four beam energies.[2]

about 200 %

titanium transmission relative to zirconia at 60 kVp[2]

about 530 %

titanium transmission relative to zirconia at 90 kVp[2]

about 5 %

X-rays passing through 4 mm of zirconia, at all four energies[2]

Because so little passed through the zirconia, raising the beam energy did not reduce its artefacts, while titanium artefacts fell as the energy rose.[2] The author limits this to the 60 to 90 kVp range tested.[2] It is one scanner and two test objects, so it shows a mechanism and does not rank every crown or implant.

## What does movement do?

During a scan the head has to stay still. A review of published sharpness data lists patient movement, with the detector and the reconstruction process, among the factors that set the effective resolution of a CBCT image. It says that once movement and scatter are counted in scans of living patients, a realistic resolution is just above 1 line pair per millimetre.[3] A line pair is one dark and one light line side by side, so that is detail of roughly half a millimetre (our arithmetic). The article on [voxel size and sharpness](https://jawviewer.com/blog/cbct-voxel-size-sharpness.md) separates the grid size from sharpness in more detail.

## How often are scans repeated for movement?

Two studies counted scans that had to be redone because of motion. They found different rates.

| Study | Patients | Repeated for motion | Where it was higher |
| --- | --- | --- | --- |
| Nardi 2017, one unit in Florence | 750 | 5.4% (41 patients) | 10.6% at age 66 and over, 5.3% at 18 and under, 3.8% at 19 to 65[4] |
| Kalabalık 2024, Turkish subpopulation | 6364 | 1.96% | Males, and children under 10[5] |

The studies differ in clinic, scanner and population, and neither explains the gap between them. Both link the repeat rate to age. Nardi and colleagues also compared three head supports: a foam headrest, a headrest with a strap and a restraint helmet. Rates were similar in people under 66, and in those over 65 the helmet clearly lowered the repeat rate.[4] The Turkish study adds that the rate was significantly higher in males than in females.[5]

> Reading these rates
>
> They count repeats in a clinic's records. They say nothing about how blurred any single scan is, or whether yours has blur at all. Ask your clinic or radiologist if you want to know.

## What can you see in the viewer?

Jaw Viewer shows a scan in slices from three directions as well as in 3D, and it does not remove artefacts. Scroll a slice from a crown towards the middle of the jaw and watch whether a streak fades. A streak that runs between two metal parts across several slices is part of the data. The same goes for a bright or dark patch touching a filling. [Window and level](https://jawviewer.com/blog/window-level-scan-too-bright.md) changes how strongly those grey bands show against bone, and the [slice views](https://jawviewer.com/blog/axial-coronal-sagittal-views-explained.md) let you check a spot from more than one direction. The [view and measure guide](https://jawviewer.com/guide/view-and-measure.md) covers the controls.

## What does it mean for measuring?

Artefacts can interfere with reading a scan, as the review above puts it.[1] A distance between two clear points of bone well away from metal is on safer ground than one that starts or ends against a crown, implant or brace. Check a point near metal in the slices before you trust the number, and talk to the clinician who ordered the scan about anything you want to rely on. Our piece on [how accurate CBCT measurements are](https://jawviewer.com/blog/how-accurate-are-cbct-measurements.md) covers the wider picture.

This is general information, not medical advice. Jaw Viewer is not a medical device. Talk to your surgeon, orthodontist or radiologist about your own case.

## Questions and answers

### Why does my CBCT scan have streaks around a crown or implant?

Dense material such as metal causes beam hardening artefacts, a common image quality problem in jaw CBCT. They come from the rebuilding process meeting the real X-ray beam, so they are in the data and not in your anatomy.

### Is a blurry CBCT scan the result of movement?

It can be. Patient movement is one of several factors, with the detector and the reconstruction, that lower the effective sharpness of a CBCT image. Your clinic or radiologist can say whether that applies to your scan.

### How often are CBCT scans repeated because of movement?

Two studies found 5.4% of 750 patients in one Italian unit and 1.96% of 6364 patients in a Turkish population. Rates were higher in children under 10 and, in the Italian study, in people aged 66 and over.

### Can I trust a measurement taken right next to metal?

Treat it with care. Artefacts can interfere with reading a scan, so a number taken against a crown, implant or brace is best checked in the slices and by the clinician who ordered the scan.

## Terms used

- **CBCT**: Cone beam computed tomography. A scanner that rotates once around the head and records a cone-shaped X-ray beam, common in dentistry and jaw surgery.
- **Beam hardening**: A scan artefact near dense material such as metal, where X-rays are filtered unevenly and dark or bright streaks appear.
- **Line pair**: One dark line and one light line side by side. Spatial resolution is often given in line pairs per millimetre.

## Sources

1. Schulze R, Heil U, Gross D et al. (2011). Artefacts in CBCT: a review. Dentomaxillofac Radiol, vol. 40, pp. 265-73. https://doi.org/10.1259/dmfr/30642039
2. Schulze R (2022). CBCT artefact-burden of zirconia-based as compared to titanium implants for different beam energies: an analytical approach. Sci Rep, vol. 12, pp. 15276. https://doi.org/10.1038/s41598-022-19379-y One zirconia implant and one titanium rod in one scanner, 60 to 90 kVp.
3. Nardi C, Taliani GG, Castellani A et al. (2017). Repetition of examination due to motion artifacts in horizontal cone beam CT: comparison among three different kinds of head support. J Int Soc Prev Community Dent, vol. 7, pp. 208-13. https://doi.org/10.4103/jispcd.JISPCD_17_17 750 patients at one radiology unit in Florence.
4. Kalabalık F, Aytuğar E, Aykanat F et al. (2024). Repetition rate of scanning due to motion artefacts in cone-beam computed tomography: a retrospective study. J Oral Maxillofac Res, vol. 15, pp. e6. https://doi.org/10.5037/jomr.2024.15106 Retrospective data of 6364 patients in a Turkish subpopulation.
5. Brüllmann D, Schulze RK (2015). Spatial resolution in CBCT machines for dental/maxillofacial applications, what do we know today?. Dentomaxillofac Radiol, vol. 44, pp. 20140204. https://doi.org/10.1259/dmfr.20140204

---