# CBCT vs medical CT: what is different

> A dental CBCT and a hospital CT both build slices from X-rays. How the image quality, grey values and dose differ, and what the studies measured.

- URL: https://jawviewer.com/blog/cbct-vs-medical-ct-differences
- Category: Understanding your scan
- Published: 2026-11-21
- Updated: 2026-11-21
- Author: Jaw Viewer
- Reading time: 4 min
- Tags: cbct, comparison, image-quality, radiation

## Key points

- In one phantom study, image contrast-to-noise ratio was 8.2 to 18.8 for four CBCT scanners and 13.6 to 20.7 for two multislice CT scanners, and low-dose CT protocols reached CBCT levels.
- In a Leuven study with the 2007 ICRP weights, effective dose was 13 to 82 microsieverts for three CBCT units and 474 to 1160 microsieverts for three multislice CT units.
- CBCT grey values vary with field size, scatter and reconstruction, so they are not Hounsfield units, and a review advised against using them quantitatively.
- CBCT is built for high-contrast structures such as bone, teeth and air spaces. Soft tissue contrast is an area where it is still being improved.

A medical CT and a dental CBCT both turn X-rays taken from many angles into a stack of slices, so a reader can look at either from three directions or as a 3D model. They differ in how the beam is shaped, how much of the head they cover, how clean the image is, what the grey numbers mean and how much radiation a scan delivers.

## How the two scanners record the image

CBCT uses a cone-shaped X-ray beam, and its image quality and dose depend on both geometric and software settings.[1] A hospital multislice CT (MSCT) is the scanner most people mean by "a CT". Both end up as a grid of voxels saved as DICOM files. Since the end of the 1990s, CBCT has become the standard three-dimensional X-ray technique in dental and jaw imaging.[2]

## Which one has the cleaner image?

One phantom study put four dental CBCT scanners and two MSCT scanners through the same image tests. The contrast-to-noise ratio, a measure of how clearly two grey levels stand apart from the grain in the image, was 8.2 to 18.8 for the CBCT scanners and 13.6 to 20.7 for the MSCT scanners. Low-dose MSCT protocols gave contrast-to-noise ratios comparable with CBCT.[3]

Sharpness was measured too. The 10% modulation transfer function, a sharpness measure, ranged from 0.1 to 0.8 per millimetre for the CBCT scanners and was 0.5 per millimetre for every MSCT protocol tested.[3]

| Phantom study | CBCT (4 scanners) | MSCT (2 scanners) |
| --- | --- | --- |
| Contrast-to-noise ratio | 8.2 to 18.8 | 13.6 to 20.7 |
| 10% modulation transfer function | 0.1 to 0.8 per mm | 0.5 per mm |

[3]

The authors concluded that CBCT scanners give adequate image quality for dentomaxillofacial examinations, and they stressed the large variation between scanners.[3] So "which is cleaner" has no single answer. The CBCT range overlaps the CT range, and settings matter as much as the scanner type.

## Can a CBCT show soft tissue?

CBCT displays high-contrast structures of the jaw region, meaning bone, teeth and air cavities, at high resolution.[4] A 2008 review lists improving soft tissue contrast among the enhancements expected in future CBCT systems.[1] Our [article on what a CBCT shows](https://jawviewer.com/blog/what-a-cbct-scan-shows.md) covers sharpness and metal in more detail.

*Figure: Axial, coronal and sagittal slices of a CBCT jaw scan beside a 3D rendering in the viewer. One public sample CBCT shown as three slice directions and as a 3D rendering. Sample scan "CBCT-MR Head (DZ-CBCT)", 3D Slicer sample data, donated for use without restrictions. Screenshot from Jaw Viewer.*

## Are the grey numbers the same?

On a medical CT the grey numbers are Hounsfield units, measured on a scale tied to water. On a CBCT they are grey values, and a 2015 review found that these can vary a great deal. It named three causes that come with the technique: the limited field size, a relatively high amount of scattered radiation, and limits of the reconstruction methods in use. The authors proposed that quantitative use of grey values on CBCT should generally be avoided at this time.[4]

The mechanism and the conversion studies have their own page: [Hounsfield units and CBCT bone density](https://jawviewer.com/blog/hounsfield-units-cbct-bone-density.md).

> One number, two meanings
>
> A value of 300 on a CT file and a value of 300 on a CBCT file are not the same quantity. In the viewer's measure tool, a CT file shows the values its own header declares. A CBCT file shows its raw grey values. The viewer does no conversion between the two.

## How different is the dose?

For the wider dose ranges, see [our dose article](https://jawviewer.com/blog/radiation-dose-cbct-vs-ct.md). A Leuven phantom study used the 2007 ICRP weights and measured effective doses of 13 to 82 microsieverts for three CBCT scanners, against 474 to 1160 microsieverts for three MSCT scanners. The CT protocols included a mandibular one with a smaller field.[5] A second study, using the older 1990 weights, found 14 to 269 microsieverts for four CBCT scanners and 350 to 742 for two MSCT scanners.[3]

Both papers show a wide spread within each scanner type. The Leuven authors stress that the choice of technique must be justified and the dose optimised.[5]

## What does the field of view change?

A CBCT scan can cover a few teeth or a larger part of the head. Why a scan stops where it does, and what that means for what you can see in it, is in [the field of view explainer](https://jawviewer.com/blog/cbct-field-of-view-explained.md). Streaks and blur have their own page as well: [metal and movement artefacts](https://jawviewer.com/blog/cbct-artefacts-streaks-near-metal-motion.md).

## Looking at both in Jaw Viewer

Jaw Viewer opens CT and CBCT DICOM files, so the same bone can be viewed from either kind of scan. Slices in three directions, 3D, and the measure tool behave the same way on both. The [view and measure guide](https://jawviewer.com/guide/view-and-measure.md) shows how, and the [open your scan guide](https://jawviewer.com/guide/open-your-scan.md) covers discs, folders and zip files.

This is general information, not medical advice. Jaw Viewer is not a medical device. Talk to your surgeon, orthodontist or radiologist about your own case.

## Questions and answers

### Is a dental CBCT a real CT scan?

Yes. Both build a stack of slices from X-rays taken at many angles. They differ in beam shape, field size, image noise, grey-value scale and dose. Studies describe CBCT as adequate for dentomaxillofacial examinations.

### Can a CBCT show soft tissue?

Poorly. CBCT shows high-contrast structures such as bone, teeth and air spaces well. A 2008 review lists better soft tissue contrast among the improvements expected in future systems.

### Are CBCT grey values the same as Hounsfield units?

No. A 2015 review found that grey values on CBCT vary because of limited field size, scatter and reconstruction, and advised against using them quantitatively at this time.

### Does a CBCT use less radiation than a medical CT?

In a Leuven phantom study, three CBCT units measured 13 to 82 microsieverts and three multislice CT units 474 to 1160 microsieverts. Doses vary a lot between machines and settings.

## Terms used

- **CBCT**: Cone beam computed tomography. A scanner that rotates once around the head and records a cone-shaped X-ray beam, common in dentistry and jaw surgery.
- **Multislice CT**: The hospital CT scanner type, MSCT for short. It records the body in thin slices as the X-ray source turns, and it is the scanner most people mean by "a CT".
- **Voxel**: A three-dimensional pixel. A scan is a grid of voxels, each with one grey value.
- **DICOM**: The file format and standard that medical scanners use for images and their details.
- **Hounsfield units**: The CT scale for how strongly a voxel blocks X-rays, with water at 0 and air near -1000.
- **Grey value**: The brightness number stored for a voxel. On a CBCT scan it is not a calibrated measure of density.

## Sources

1. Scarfe WC, Farman AG (2008). What is cone-beam CT and how does it work?. Dent Clin North Am, vol. 52, pp. 707-30. https://doi.org/10.1016/j.cden.2008.05.005
2. Suomalainen A, Kiljunen T, Käser Y et al. (2009). Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners. Dentomaxillofac Radiol, vol. 38, pp. 367-78. https://doi.org/10.1259/dmfr/15779208 Phantom study, four CBCT and two multislice CT scanners. An erratum was published in 2009.
3. Loubele M, Bogaerts R, Van Dijck E et al. (2009). Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications. Eur J Radiol, vol. 71, pp. 461-8. https://doi.org/10.1016/j.ejrad.2008.06.002 Phantom study, three CBCT and three multislice CT scanners.
4. Pauwels R, Jacobs R, Singer SR et al. (2015). CBCT-based bone quality assessment: are Hounsfield units applicable?. Dentomaxillofac Radiol, vol. 44, pp. 20140238. https://doi.org/10.1259/dmfr.20140238
5. Schulze RKW, Drage NA (2020). Cone-beam computed tomography and its applications in dental and maxillofacial radiology. Clin Radiol, vol. 75, pp. 647-57. https://doi.org/10.1016/j.crad.2020.04.006

---