# What a CBCT scan shows, and what it cannot tell you

> How sharp a CBCT image really is, why its grey values are not bone density, what metal does to the picture, and what that means for measurements.

- URL: https://jawviewer.com/blog/what-a-cbct-scan-shows
- Category: Understanding your scan
- Published: 2026-11-20
- Updated: 2026-11-20
- Author: Jaw Viewer
- Reading time: 4 min
- Tags: cbct, image-quality, measurements

## Key points

- Published CBCT resolution measurements range from 0.6 to 2.8 line pairs per millimetre, and a review expects just above 1 in real patient scans, which is features about half a millimetre wide.
- CBCT grey values cannot be reliably converted to Hounsfield units, so they are not a measure of bone density.
- Metal such as fillings and braces causes beam hardening streaks that can hide detail next to it.

You can scroll through a CBCT scan, turn it and read measurements to a tenth of a millimetre. Those decimal places can suggest more precision than the image supports. Sharpness, grey values and artefacts all affect what you can read from it.

## What it is good for

CBCT came into dental and jaw imaging at the end of the 1990s and has since become the standard three-dimensional X-ray technique in that field, with a wide range of uses.[1] It builds a stack of slices from one turn around the head, which is why the same scan can be shown as slices from three directions or as a 3D model. Our article on [radiation dose](https://jawviewer.com/blog/radiation-dose-cbct-vs-ct.md) covers what a scan costs in dose.

## How sharp it really is

A scan's voxel size, such as 0.25 mm, says how big each voxel is. It does not say how small a detail you can see. Sharpness is measured as spatial resolution, often in line pairs per millimetre (lp/mm). A review of published CBCT measurements found values from 0.6 to 2.8 lp/mm, with a median of about 1.7 lp/mm by visual assessment. It also found that real scans of living patients, where the person moves a little and scatter and other effects degrade the image, can realistically expect just above 1 lp/mm. That is roughly one order of magnitude lower than an intraoral X-ray.[2]

One line pair per millimetre means a dark line and a light line side by side fit in one millimetre, so the thinnest line you can tell apart is about half a millimetre wide (our arithmetic). It follows (our inference, not a finding of the review) that very thin structures, such as a sinus wall, can look blurred or incomplete because they are thinner than the scan can resolve. The same review lists what limits sharpness: the detector, the reconstruction process and patient movement.[2]

## Grey values are not bone density

Each voxel holds a grey value. On a medical CT those numbers are Hounsfield units (HU), a calibrated scale. On a CBCT the numbers look similar but are a different thing. A systematic review asked whether CBCT grey values can be converted into Hounsfield units. Only three studies qualified, and the review concluded that they cannot be converted, because clinical studies showing it works are missing. It adds that doing it properly would need calibration of the machine, a prediction model and a standard formula, not a number read off the software.[3]

Another review and meta-analysis did find that CBCT grey values and CT Hounsfield units are positively correlated, with a moderate correlation of r = 0.669 (95% confidence interval 0.388 to 0.836) in the two studies it could pool.[4] The two measures tend to move in the same direction, but one cannot stand in for the other.

> What this means for a patient
>
> A viewer will show you a number when you point at a spot on a CBCT scan. Treat it as that scanner's grey value, good for comparing places within one scan. Do not read it as bone density, and do not compare it with a number from a different scanner.

## What metal does to the picture

Fillings, crowns, braces and implants are dense. Beam hardening artefacts caused by metal are described as a common quality issue in jaw CBCT images.[5] In a study comparing zirconia and titanium implants at four beam energies, artefacts from titanium decreased as the beam energy rose, while those from zirconia ceramic did not.[5] So the area around metal is where to look hardest, in the slices as well as the 3D rendering.

## What this means for measurements

Distances between two clear bony points are what a CBCT scan is used for. The limits above matter most for small things: thin walls, small gaps, and anything next to metal. If a measurement depends on a structure that is a millimetre thick or less, or sits against a crown or implant, treat the number as approximate and look at the slices as well as the 3D model.

This is general information, not medical advice. Jaw Viewer is not a medical device. Talk to your surgeon, orthodontist or radiologist about your own case.

## Questions and answers

### How small a detail can a CBCT scan show?

A review of published measurements found values between 0.6 and 2.8 line pairs per millimetre, and expects just above 1 line pair per millimetre in real scans of patients. One line pair per millimetre means features about half a millimetre wide.

### Can I read bone density from the grey values?

Not reliably. A 2022 systematic review concluded that CBCT grey values cannot be converted to Hounsfield units of a medical CT. A meta-analysis found a moderate correlation between the two.

### Why are there streaks around fillings and braces?

Dense material such as metal causes beam hardening artefacts, a common image quality problem in jaw CBCT.

## Terms used

- **CBCT**: Cone beam computed tomography. A scanner that rotates once around the head and records a cone-shaped X-ray beam, common in dentistry and jaw surgery.
- **Voxel**: A three-dimensional pixel. A scan is a grid of voxels, each with one grey value.
- **Line pair**: One dark line and one light line side by side. Spatial resolution is often given in line pairs per millimetre.
- **Hounsfield units**: The CT scale for how strongly a voxel blocks X-rays, with water at 0 and air near -1000.
- **Grey value**: The brightness number stored for a voxel. On a CBCT scan it is not a calibrated measure of density.

## Sources

1. Schulze RKW, Drage NA (2020). Cone-beam computed tomography and its applications in dental and maxillofacial radiology. Clin Radiol, vol. 75, pp. 647-57. https://doi.org/10.1016/j.crad.2020.04.006
2. Brüllmann D, Schulze RK (2015). Spatial resolution in CBCT machines for dental/maxillofacial applications, what do we know today?. Dentomaxillofac Radiol, vol. 44, pp. 20140204. https://doi.org/10.1259/dmfr.20140204
3. Eguren M, Holguin A, Diaz K et al. (2022). Can gray values be converted to Hounsfield units? A systematic review. Dentomaxillofac Radiol, vol. 51, pp. 20210140. https://doi.org/10.1259/dmfr.20210140 Three studies met the inclusion criteria.
4. Selvaraj A, Jain RK, Nagi R et al. (2022). Correlation between gray values of cone-beam computed tomograms and Hounsfield units of computed tomograms: a systematic review and meta-analysis. Imaging Sci Dent, vol. 52, pp. 133-40. https://doi.org/10.5624/isd.20210274 14 studies in the review, 2 in the meta-analysis.
5. Schulze R (2022). CBCT artefact-burden of zirconia-based as compared to titanium implants for different beam energies: an analytical approach. Sci Rep, vol. 12, pp. 15276. https://doi.org/10.1038/s41598-022-19379-y

---